100 Watt
Universal Input For Medical
Medical AC-DC power supplies for medical applications
Our 100 watt universal input AC-DC converter for medical applications features active power factor correction, a built-in EMI Class A filter and a low leakage current. This model also has an open frame package of 2.00" x 3.00" x 1.16" and meets the following medical safety approvals: UL, CE, ES60601-1, EN60601-1 and ANSI/AAMI.

INPUT
- Voltage Range
- AC Input: 85 - 264 VAC
- DC Input: 120 - 370 VDC
- Input Frequency
- AC Input: 47 - 63 Hz
- Input Current
- 115 VAC and Full Load: 1.15 A
- 230 VAC and Full Load: 0.55 A
- No Load Input Power
- 230 VAC: 0.3 Watts
- Leakage Current
- 264 VAC: 75µA
- Power Factor: 0.95
- Start up time: 1000 ms
- Rise time: 20 ms
- Hold up time
- 115 VAC and Full Load: 16 ms
- Input inrush current
- 230 VAC: 60 A
- Input protection
- Internal fuse inline and neutral: T3,15A/250 VAC
OUTPUT
- Output Power
- Open frame: 100 Watts
- Initial set voltage accuracy
- 230 VAC and full Load: ±1.0%
- Line Regulation
- Low Line to High Line at Full Load: ±0.2%
- Load Regulation
- No Load to Full Load: ±0.5%
- 10 - 90% Load: ±0.4%
- Voltage Adjustability: ±10%
- Minimum Load: 0%
- Ripple and Noise (Measured by 20 MHz band width)
- With a 10µF/25V 1206 X7R MLCC
- 12 Vout: 120 mVp-p
- With a 10µF/25V 1206 X7R MLCC
- 15 Vout: 150 mVp-p
- With a 1µF/50V 1206 X7R MLCC
- 24 Vout: 160 mVp-p
- With a 1µF/50V 1206 X7R MLCC
- 28 Vout: 180 mVp-p
- With a 1µF/50V 1206 X7R MLCC
- 36 Vout: 190 mVp-p
- With a 0.1µF/100V 1206 X7R MLCC
- 48 Vout: 340 mVp-p
- With a 10µF/25V 1206 X7R MLCC
- Temperature Coefficient: ±0.2%°C
- Transient Response
- Load step from 50 - 75% change at 2.5A/µs
- Peak deviation: 3% Vout
- Recovery time: 500µs
- Load step from 50 - 75% change at 2.5A/µs
- Over Voltage Protection
- Percentage of Vout (nom), Latch mode: 135%
- Over Load Protection
- Percentage of lout rated, hiccup mode: 150%
- 12 Vout, 15 Vout
- Percentage of lout rated, hiccup mode: 150%
- Short Circuit Protection: continuous, automatic recovery
GENERAL
- Efficiency: 91 - 92%
- Isolation Voltage 1 minute (2MOPP insulation)
- Input to Output: 4000 VAC
- Input (Output) to F.G: 1500 VAC
- Isolation Resistance
- 500 Vdc: 0.1GΩ
- Switching Frequency: 60 kHz
- Approvals and Standard: ANSI/AAMI, ES60601-1, IEC6060-1 and EN60601-1
- Weight
- MUIO: 156 g (5.50 oz)
- MUIT: 194 g (6.84 oz)
- MUI: 210 g (7.41 oz)
- MUID: 232 g (8.18 oz)
- MTBF
- MIL-HDBK-217F Ta25°C, Full load: 7.903 x 10^5 hrs
ENVIRONMENTAL
- Operating Ambient Temperature: -25 - 80°C with derating
- Storage Temperature Range: -40 - 85°C
- Operating Altitude: 5000 m
- Thermal Shock: MIL-STD-810F
- Shock: IEC68-2-27
- Vibration: IEC68-2-6
- Relative Humidity
- Non-condensing: 5 - 95% RH
EMC CHARACTERISTICS
- EMI
- EN55011, EN55022 and FCC Part 18: Class A, Class B
- Harmonic Currents
- EN61000-3-2, Full Load: Class A and D
- Voltage Flicker: EN61000-3-3
- ESD
- EN61000-4-2, Perf. Criteria A
- Air ±8KV
- Contact ±6KV
- EN61000-4-2, Perf. Criteria A
- Radiated Immunity
- EN61000-4-3: 20 V/m, Perf. Criteria A
- Fast Transient
- EN61000-4-4: ±2KV, Perf. Criteria A
- Surge
- EN61000-4-5, Perf. Criteria A
- DM ±1KV
- CM ±2KV
- EN61000-4-5, Perf. Criteria A
- Conducted Immunity
- EN61000-4-6: 20 Vrms, Perf. Criteria A
- Dip and Interruptions
- EN60601-1-2
- 230 VAC 50 Hz
- 30%: 500 mS, Perf. Criteria A
- 60%: 100 mS, Perf. Criteria A
- >95%: 10 mS, Perf. Criteria A
- >95%: 5000 mS, Perf. Criteria B
- 100 VAC 50 Hz
- 30%: 500 mS, Perf. Criteria A
- 60%: 100 mS, Perf. Criteria B
- >95%: 10 mS, Perf. Criteria A
- >95%: 5000 mS, Perf. Criteria B
- 230 VAC 50 Hz
- EN60601-1-2
Input Range (Vac) | Output Voltage(Vdc) | Output Current @ Natural Convection 50ºC Ta (A) | Input Power @ No Load (W) | Efficiency (%) | Model Number* |
85-264 | 12 | 8.34 | 0.3 | 91 | MUI100-12SB |
15 | 6.67 | 0.3 | 92 | MUI100-15SB | |
24 | 4.17 | 0.3 | 92 | MUI100-24SB | |
28 | 3.58 | 0.3 | 92 | MUI100-28SB | |
36 | 2.78 | 0.3 | 91 | MUI100-36SB | |
48 | 2.09 | 0.3 | 91 | MUI100-48SB |