Contact us | +1 973.345.5885 |

120 Watt, Universal Input, PFMUIE120 Medical Power Supplies

120 Watt

Universal Input


PFMUIE120 Medical Power Supplies

PFMUIE120 Series enclosed switching power supplies are well-suited for medical applications. Delivering 120 watts of output with 10-CFM forced air and 85 watts with free air convection, the units provide 4,000V AC input to output 2xMOPP insulation for patient safety, meet the UL/IEC/EN 60601-1 edition 3.1 medical safety standard and EN55011 EMI standards, and they are suitable for body floating applications when used with an appropriate system. The PFMUIE120 measures 3.15 by 2.35 by 1.7 inches and offers a 90 to 264V AC universal input range, efficiency to 91 percent, 0.3-W no-load input power and a PFC function of >0.9.

PFMUIE120

 

OUTPUT WATTAGE

  • Max Output Wattage (with 10CFM FAN) (W)
    • 120 W
  • Max Output Wattage (Free air Convection) (W)
    • 85 W

INPUT

  • Voltage
    • 90-264 VAC
  • Frequency (Hz)
    • 47-63 Hz
  • Current (Full load)
    • < 2.0 A max. (115 VAC)/< 1.0 A max. (230 VAC)
  • Inrush Current (<2ms)
    • < 40 A max. (115 VAC)/< 70 A max. (230 VAC)
  • Leakage Current
    • < 0.1mA / 264 VAC (Touch Current)
  • Power Factor (at 230 VAC)
    • PF>0.9 at Full Load
  • No Load
    • < 0.3W (115 / 230 VAC)

OUTPUT

  • Voltage (V.DC.)
    • PFMUIE120-12S: 12V
    • PFMUIE120-24S: 24V
    • PFMUIE120-48S: 48V
  • Voltage Adj. Range (V.DC): ±10% Output Voltage
  • Voltage Accuracy: ±2%
  • Current (with 10CFM FAN) (A) (max.)
    • PFMUIE120-12S: 10
    • PFMUIE120-24S: 5
    • PFMUIE120-48S: 2.5
  • Current (Free air Convection) (A) (max.)
    • PFMUIE120-12S: 7.083
    • PFMUIE120-24S: 3.542
    • PFMUIE120-48S: 1.771
  • Line Regulation: ±1%
  • Load Regulation (10-100%): ±1%
  • Minimum Load: 0%
  • Maximum Capacitive Load
    • PFMUIE120-12S: 3000μF
    • PFMUIE120-24S: 1500μF
    • PFMUIE120-48S: 500μF
  • Ripple & Noise (max.)
    • PFMUIE120-12S: 160mV
    • PFMUIE120-24S: 1% Vout
    • PFMUIE120-48S: 1% Vout
  • Efficiency (at 230 VAC)
    • PFMUIE120-12S: 90%
    • PFMUIE120-24S: 90%
    • PFMUIE120-48S: 91%
  • Hold-up Time (at 115 VAC): 10 ms min.

PROTECTION

  • Over Power Protection
    • Auto recovery, Hiccup mode
  • Over Voltage Protection
    • Latch off
  • Overt Temperature Protection
    • Latch off
  • Short Circuit Protection
    • Protection level 1 (nominal): Continuous, Auto recovery
    • Protection level 2 (instantaneous high current): Latch

ISOLATION

  • Input-Output
    • 4000VAC or 5656VDC
  • Input-FG
    • 2000VAC or 2828VDC
  • Output-FG
    • 1500VAC or 2121VDC

ENVIRONMENT

  • Operating Temperature
    • -30°C...+70°C (with derating)
  • Storage Temperature
    • -30°C...+85°C
  • Temperature Coefficient
    • ±0.05%/°C
  • Altitude During Operation
    • 5000m
  • Humidity
    • 20~90% RH
  • Atmospheric Pressure
    • 56 kPa to 106 kPa
  • MTBF
    • >250,000 h @ 25°C (MIL-HDBK-217F, Notice 1)
  • Vibration
    • IEC60068-2-6 (10~500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes)
  • Shock
    • IEC60068-2-27

PHYSICAL

  • Dimensions (L x W x H)
    • 3.15 x 2.35 x 1.7 Inches ( 80.0 x 59.7 x 43.2 mm ) Tolerance ±0.5 mm
  • Weight
    • 258 g

SAFETY

  • Approval
    • Others: UL / IEC / EN 60601 3.1rd Edition

EMC

  • Conducted EMI
    • EN55011 Conducted Class B
  • Radiated EMI
    • EN55011 Class I class B / Class II class A
  • EMS
    • EN60601-1-2 4th edition