240 Watt
Universal Input
PFMUIE240 Medical Power Supplies
The PFMUIE240 Series of 240-watt switching power supplies comes in a 4.1 by 2.44 by 1.99-inch enclosed package and accepts a 90 to 264V AC universal input range. Providing 4,000V AC input to output 2xMOPP insulation for patient safety, these medical power supplies meet the UL/IEC/EN 60601-1 edition 3.1 safety standard and EN55011 EMI standards, and they are suitable for body floating applications when used with an appropriate system. These supplies offer efficiency to 93 percent, 0.5-W no-load input power and a PFC function of >0.9.

OUTPUT WATTAGE
- Max Output Wattage (with 10CFM FAN) (W)
- 240 W
INPUT
- Voltage
- 90-264 VAC
- Frequency (Hz)
- 47-63 Hz
- Current (Full load)
- < 3.0Amax. (115VAC)/< 1.5Amax. (230VAC)
- Inrush Current (<2ms)
- < 45Amax. (115VAC)/< 90Amax. (230VAC)
- Leakage Current
- < 0.1mA / 264 VAC (Touch Current)
- Power Factor
- PF>0.9 at Full Load
OUTPUT
- Voltage (V.DC.)
- PFMUIE240-12S: 12V
- PFMUIE240-15S: 15V
- PFMUIE240-24S: 24V
- PFMUIE240-48S: 48V
- Voltage Adj. Range (V.DC): ±4% Output Voltage
- Voltage Accuracy: ±2%
- Current (A) (max.)
- PFMUIE240-12S: 20
- PFMUIE240-15S: 16
- PFMUIE240-24S: 10
- PFMUIE240-48S: 5
- Line Regulation: ±1%
- Load Regulation (10-100%) (typ.): ±1%
- Minimum Load: 0%
- Maximum Capacitive Load
- PFMUIE240-12S: 8000μF
- PFMUIE240-15S: 2000μF
- PFMUIE240-24S: 3000μF
- PFMUIE240-48S: 470μF
- Ripple & Noise (max.): 1% Vout
- Efficiency (at 230 VAC)
- PFMUIE240-12S: 92%
- PFMUIE240-15S: 92%
- PFMUIE240-24S: 92.5%
- PFMUIE240-48S: 93%
- Hold-up Time (at 115 VAC): 10 ms min.
PROTECTION
- Over Power Protection
- Auto recovery, Hiccup mode
- Over Voltage Protection
- Auto recovery
- Overt Temperature Protection
- Auto recovery
- Short Circuit Protection
- Protection level 1 (nominal): Continuous, Auto recovery
- Protection level 2 (instantaneous high current): Latch
ISOLATION
- Input-Output
- 4000VAC or 5656VDC
- Input-PE
- 2000VAC or 2828VDC
- Output-PE
- 1500VAC or 2121VDC
ENVIRONMENT
- Operating Temperature
- -30°C...+70°C (with derating)
- Storage Temperature
- -30°C...+85°C
- Temperature Coefficient
- ±0.05%/°C
- Altitude During Operation
- 5000m
- Humidity
- 20~90% RH
- Atmospheric Pressure
- 56 kPa to 106 kPa
- MTBF
- >250,000 h @ 25°C (MIL-HDBK-217F, Notice 1)
- Vibration
- IEC60068-2-6 (10~500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes)
- Shock
- IEC60068-2-27
PHYSICAL
- Dimensions (L x W x H)
- 4.1 x 2.44 x 1.99 Inches ( 104.0 x 62.0 x 50.5 mm ) Tolerance ±0.5 mm
- Weight
- 328 g
SAFETY
- Approval
- Others: UL / IEC / EN 60601 3.1rd Edition & UL / IEC / EN 60950 AM2
- Approval / Meet
- 15S: UL / IEC / EN 60601 3.1rd Edition / UL / IEC / EN 60950 AM2 (meet)
EMC
- Conducted EMI
- EN55011 Conducted Class B
- Radiated EMI
- EN55011 Class I class B / Class II class A
- EMS
- EN60601-1-2 4th edition