40 Watt
Dual and Triple Output
Efficient AC-DC power supplies for medical applications
With a universal input range and high efficiency up to 90%, our 40-watt MUI40 dual and triple output series of AC-DC converters for medical applications has a low leakage current under 75µA and a built-in EMI class B filter. This model also features an operating altitude of 5000M, 0.15-watt low standby power consumption and a 4000 VAC input to output 2 MOPP insulation. In addition, this model meets ANSI/AAMI ES60601-1, EN60601-1 and IEC60601-1 3rd Edition safety standards, is CE Marked: IEC60601-1-2 ED4.0 and is compliant to RoHS II & Reach.

INPUT
- Voltage Range:
- AC Input: 85 - 264VAC
- DC Input: 120 - 370VDC
- Input Frequency: AC Input: 47 - 63Hz
- Input Current:
- 100 VAC: 1.05A
- 240 VAC: 0.55A
- No Load Input Power: 230 Vac: 0.15 Watts
- Leakage Current: 264 Vac: 75µA
- Start-up Time: 1000ms
- Rise time: 20ms
- Hold-up time: 115 Vac and Full Load: 25ms
- Input Inrush Current: 230 Vac: 60A
- Input Protection: Internal fuse in line and neutral: T3.15A/250Vac
OUTPUT
- Output Power: Pout1+Pout2+Pout3: 40W
- Initial Set Voltage Accuracy:
- 230Vac and Full Load, Vout 1: -1 - 1%
- Vout 2, Vout 3: -2 -2%
- Line Regulation: Low Line to High Line at Full Load: -0.2 - 0.2%
- Load Regulation:
- No Load to Full Load, Vout 1: -0.5 - 0.5%
- Vout 3: -0.7 - 0.7%
- No Load to Full Load, Vout 2: -1.5 - 1.5%
- 0.1W Load to Full Load, Vout 2: -0.7 - 0.7
- Cross Regulation: Asymmetrical load 25%/100% FL: -1.5 - 1.5%
- Voltage Adjustability: Vout 1: -10 - 10%
- Minimum Load:
- 0%
- Vout 3 is full load, Vout 1 + Vout 2: 0.5%
- Ripple and Noise (Measured by 20MHz bandwidth)
- With a 10μF/25V 1206 X7R MLCC, Vout 1, 5V: 100mVp-p
- With a 1μF/50V 1206 X7R MLCC, Vout 1, 12V: 120mVp-p
- With a 1μF/50V 1206 X7R MLCC, Vout 1, 15V: 150mVp-p
- With a 1μF/50V 1206 X7R MLCC, Vout 1, 24V: 240mVp-p
- With a 1μF/50V 1206 X7R MLCC, Vout 1, 28V: 280mVp-p
- With a 10μF/25V 1206 X7R MLCC, Vout 2, All: 100mVp-p
- With a 10μF/25V 1206 X7R MLCC, Vout 3, 5V: 100mVp-p
- With a 10μF/25V 1206 X7R MLCC, 12V: 120mVp-p
- With a 10μF/25V 1206 X7R MLCC, 15V: 150mVp-p
- Temperature Coefficient: -0.01 - 0.02%/ºC
- Transient Response Peak Deviation: Load step from 50-75% change at 2.5A/μs, Vout 1: 3% Vout
- Transient Response Recovery Time: 600μs
- Over Voltage Protection: % of Vout(nom); Latch mode, Vout 1, 125 - 140
- Over Power Protection: % of nominal output power; Hiccup mode, Pout 1 + Pout 2: 145%
- Short Circuit Protection: Continuous, automatic recovery
GENERAL
- Isolation Voltage (1 minute, 2MOPP insulation):
- Input to Output: 4000Vac
- Input (Output) to F.G.: 2500Vac
- Isolation Resistance: 500Vdc, 0.1GΩ
- Switching Frequency:
- 230Vac, Vout 1, 5V: 70kHz
- Others: 115kHz
- Vout 2: 750kHz
- Vout 3: 510kHz
ENVIRONMENTAL
- Operating Ambient Temperature: With derating: -40 - 85ºC
- Storage Temperature Range: -40 - 85ºC
- Operating Altitude: 5000m
- Shock: IEC60068-2-27
- Vibration: IEC60068-2-6
- Relative Humidity: Non-condensing, 5% to 95% RH
PHYSICAL
- Design Meet Safety Standard: ANSI/AAMI, ES60601-1, IEC60601-1, EN60601-1
- Weight:
- 150g (5.29oz): MUIO
- 198g (6.98oz): MUIT
- 216g (7.62oz): MUI
- 238g (8.40oz): MUID
- Dimensions: 2” × 3.5” (50.8 mm × 88.9 mm)
- MTBF: 1.716 × 106 hrs , MIL-HDBK-217F, Full load
EMC SPECIFICATIONS
- EMI: EN55011, EN55022 and FCC Part 18
- Conducted: Class B
- Radiated: Class B
- Harmonic Currents: EN61000-3-2, Full load, Class A
- Voltage Flicker: EN61000-3-3
- ESD: EN61000-4-2, Air ±15kV and Contact ±8kV, Perf. Criteria A
- Radiated Immunity: EN61000-4-3, 20V/m, Perf. Criteria A
- Fast Transient: EN61000-4-4, ±2kV, Perf. Criteria A
- Surge: EN61000-4-5, DM ±1kV, Perf. Criteria A
- Conducted Immunity: EN61000-4-6, 20 Vr.m.s, Perf. Criteria A
- Power Frequency Magnetic Field: EN61000-4-8, 30 A/M, Perf. Criteria A
- Dip and Interruptions:
- EN60601-1-2, 0% of 240 Vac, 0.5 cycle, Perf. Criteria A
- EN61000-4-11, 0% of 240 Vac, 1 cycle, Perf. Criteria A
- EN61000-4-11, 70% of 240 Vac, 25/30 cycle, Perf. Criteria A
- EN61000-4-11, 0% of 240 Vac, Single phase, Perf. Criteria B
- EN61000-4-11, 0% of 100 Vac, 0.5 cycle, Perf. Criteria A
- EN61000-4-11, 0% of 100 Vac, 1 cycle, Perf. Criteria B
- EN61000-4-11, 70% of 100 Vac, 25/30 cycle, Perf. Criteria A
- EN61000-4-11, 0% of 100 Vac, Single phase, Perf. Criteria B
Input Voltage Range (Vac) |
Output 1 | Output 2 | Output 3 | Max. Output Power W |
Efficiency % |
Model Number* | |||||
Voltage Vdc |
Current (Normal) A |
Max. Current A |
Voltage Vdc |
Current (Normal) A |
Max. Current A |
Voltage Vdc |
Max. Current A |
||||
85-264 | 5 | 5 | 8 | 3.3 | 4 | 6 | 40 | 89.5 | MUI40-5S33SB | ||
12 | 2.1 | 3.34 | 5 | 4 | 6 | 40 | 89 | MUI40-1212S5SB | |||
12 | 2.1 | 3.34 | 3.3 | 4 | 6 | 40 | 90 | MUI40-12S33SB | |||
15 | 1.7 | 2.67 | 5 | 4 | 6 | 40 | 88.5 | MUI40-15S5SB | |||
24 | 1.05 | 1.67 | 5 | 4 | 6 | 40 | 86 | MUI40-24S5SB | |||
28 | 0.72 | 1.43 | 5 | 4 | 6 | 40 | 85.5 | MUI40-28S5SB | |||
5 | 5 | 8 | 3.3 | 4 | 6 | -5 | 0.5 | 40 | 89 | MUI40-5S-33B | |
5 | 5 | 8 | 3.3 | 4 | 6 | 12 | 0.5 | 40 | 89 | MUI40-5S33S12SB | |
5 | 5 | 8 | 3.3 | 4 | 6 | -12 | 0.5 | 40 | 89 | MUI40-5S33SN12SB | |
12 | 2.1 | 3.34 | 5 | 4 | 6 | -5 | 0.5 | 40 | 88.5 | MUI40-12S-5B | |
12 | 2.1 | 3.34 | 5 | 4 | 6 | -12 | 0.5 | 40 | 88 | MUI40-5S-12B | |
12 | 2.1 | 3.34 | 3.3 | 4 | 6 | 5 | 0.5 | 40 | 88.5 | MUI40-12S33S5SB | |
12 | 2.1 | 3.34 | 3.3 | 4 | 6 | -12 | 0.5 | 40 | 88 | MUI40-33S-12B | |
15 | 1.7 | 2.67 | 5 | 4 | 6 | -15 | 0.5 | 40 | 88 | MUI40-5S-15B | |
24 | 1.05 | 1.67 | 5 | 4 | 6 | 12 | 0.5 | 40 | 86 | MUI40-24S5S12SB | |
24 | 1.05 | 1.67 | 5 | 4 | 6 | -12 | 0.5 | 40 | 86 | MUI40-24S5SN12SB |