Contact us | +1 973.345.5885 |

500 Watt, Dual Inputs, PFMUIU500 Medical Power Supplies

500 Watt

Dual Inputs


PFMUIU500 Medical Power Supplies

PFMUIU500 Series 500-watt switching power supplies feature an ultra-small 5.5 by 3.25 by 1.6-inch U-frame design for space-constrained medical applications. The units accept inputs of 90 to 264V AC and 127 to 370V DC, and they deliver 500 watts of output power with a 30-CFM fan and 200 watts with free air convection. Providing 4,000V AC input to output 2xMOPP insulation for patient safety, these medical power supplies meet the UL/IEC/EN 60601-1 edition 3.1 safety standard, and they offer high efficiency to 93 percent, standby of 5V at 1A with fan and 0.4A without fan, remote ON/OFF, a PFC function of >0.94 and optional current share function on most models. PFMUIU500 power supplies also come with a 12-V, 0.3-A auxiliary output.

PFMUIU500

 

OUTPUT WATTAGE

  • Max Output Wattage (W)
    • 500 W (30CFM FAN)
  • Max Output Wattage (W)
    • Others: 190 W (115 VAC) / 200 W (230 VAC)
    • 15S: 170 W (115 VAC) / 180 W (230 VAC)

INPUT

  • Voltage
    • 90-264 VAC or 127-370 VDC
  • Frequency (Hz)
    • 47-63 Hz
  • Current (Full load)
    • < 6.3 A max. (115 VAC)/< 3.15 A max. (230 VAC)
  • Inrush Current (<2ms) (Clod Start)
    • < 40 A max. (115 VAC)/< 80 A max. (230 VAC)
  • Leakage Current
    • < 0.1 mA max. (Input-Output)
  • Power Factor (at 230 VAC)
    • PF>0.94 at Full Load

OUTPUT

  • Voltage (V.DC.)
    • PFMUIU500-12S: 12V
    • PFMUIU500-15S: 15V
    • PFMUIU500-24S: 24V
    • PFMUIU500-48S: 48V
  • Voltage Accuracy: ±2%
  • Voltage Adj. Range (V.DC): ±4% Output Voltage
  • Current (with 30CFM FAN) (A) max
    • PFMUIU500-12S: 41.5
    • PFMUIU500-15S: 33.3
    • PFMUIU500-24S: 20.8
    • PFMUIU500-48S: 10.41
  • Current (Free air Convection) (A) max
    • PFMUIU500-12S: 15.83 at 115 VAC; 16.6 at 230 VAC
    • PFMUIU500-15S: 11.33 at 115 VAC; 12 at 230 VAC
    • PFMUIU500-24S: 7.91 at 115 VAC; 8.33 at 230 VAC
    • PFMUIU500-48S: 3.96 at 115 VAC; 4.17 at 230 VAC
  • Line Regulation (115-264 VAC): ±0.5%
  • Load Regulation (10-100%) (typ.): ±1%
  • Minimum Load: 3%
  • Maximum Capacitive Load
    • PFMUIU500-12S: 5,000μF
    • PFMUIU500-15S: 3,750μF
    • PFMUIU500-24S: 2,500μF
    • PFMUIU500-48S: 1,250μF
  • Ripple & Noise (max.)
    • PFMUIU500-12S: 160mV
    • PFMUIU500-15S: 160mV
    • PFMUIU500-24S: 240mV
    • PFMUIU500-48S: 480mV
  • Efficiency (at 230 VAC)
    • PFMUIU500-12S: 90.5%
    • PFMUIU500-15S: 90.5%
    • PFMUIU500-24S: 92%
    • PFMUIU500-48S: 93%
  • Hold-up Time (at 115 VAC): 8 ms min.

PROTECTION

  • Over Power Protection
    • Auto recovery
  • Over Voltage Protection
    • Auto recovery
  • Overt Temperature Protection
    • Auto recovery
  • Short Circuit Protection
    • Protection level 1 (nominal): Continuous, Auto recovery
    • Protection level 2 (instantaneous high current): Latch

ISOLATION

  • Input-Output (V.AC)
    • 4000VAC or 5656VDC
  • Input-PE (V.AC)
    • 2000VAC
  • Output-PE
    • 1500VAC

ENVIRONMENT

  • Operating Temperature
    • -30°C...+70°C (with derating)
  • Storage Temperature
    • -35°C...+85°C
  • Temperature Coefficient
    • ±0.03%/°C (0~50°C)
    • ±0.06%/°C (-30~0°C)
  • Altitude During Operation
    • 5000m
  • Humidity
    • 95% RH
  • Atmospheric Pressure
    • 56 kPa to 106 kPa
  • MTBF
    • >160,000 h @ 25°C (MIL-HDBK-217F)
  • Vibration
    • IEC60068-2-6 (10~500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes)
  • Shock
    • IEC60068-2-27

PHYSICAL

  • Dimensions (L x W x H)
    • 5.5 x 3.25 x 1.6 Inches ( 139.7 x 82.55 x 40.6 mm ) Tolerance ±0.5 mm
  • Weight
    • 580 g
  • Cooling Method
    • Free convection / 30 CFM FAN

SAFETY

  • Approval
    • Others: UL / IEC / EN 60601 3.1rd Edition & UL / IEC / EN 60950 AM2
    • 15S: UL / IEC / EN 60601 3.1rd Edition

EMC

  • Conducted and Radiated EMI
    • EN55011 / conducted class B, Radiated Class A
  • EMS
    • EN60601-1-2 4th edition