500 Watt
Universal Input
PFUIE500 Power Supplies
PFUIE500 Series 500-watt power supplies offer 90 to 264V AC universal input and come in a 5.11 by 3.25 by 2.42-inch enclosed package for medical applications. Providing 4,000V AC input to output isolation, the units meet the UL/IEC/EN 62368-1 safety standard, and they offer high efficiency to 91.5 percent, standby of 5V at 1A, >0.94 active PFC at full load and an operating altitude of 5,000m. Safeguards include overpower, overvoltage, over temperature and short-circuit protections. The operating temperature range is -30°C to +70°C.

OUTPUT WATTAGE
- Max Output Wattage (W): 500 W
INPUT
- Voltage
- 90-264 VAC or 127-370 VDC
- Frequency (Hz)
- 47-63 Hz
- Current (Full load)
- < 6.3 A max. (115 VAC)/< 3.15 A max. (230 VAC)
- Inrush Current (<2ms) (Clod Start)
- < 40 A max. (115 VAC)/< 80 A max. (230 VAC)
- Power Factor (at 230 VAC)
- PF>0.94 at Full Load
OUTPUT
- Voltage (V.DC.)
- PFUIE500-12S: 12V
- PFUIE500-24S: 24V
- PFUIE500-48S: 48V
- Voltage Adj. Range (V.DC): ±5% Output Voltage
- Voltage Accuracy: ±2%
- Current (with 30CFM FAN) (A) max
- PFUIE500-12S: 41.6
- PFUIE500-24S: 20.8
- PFUIE500-48S: 10.41
- Line Regulation (100-264 VAC): ±1%
- Load Regulation (10-100%) (typ.): ±1%
- Minimum Load: 1%
- Maximum Capacitive Load
- PFUIE500-12S: 5,000μF
- PFUIE500-24S: 2,500μF
- PFUIE500-48S: 1,250μF
- Ripple & Noise (typ.)
- PFUIE500-12S: 160mV
- PFUIE500-24S: 240mV
- PFUIE500-48S: 480mV
- Efficiency (at 230 VAC)
- PFUIE500-12S: 90%
- PFUIE500-24S: 90.5%
- PFUIE500-48S: 91.5%
- Hold-up Time (at 115 VAC): 8 ms min.
PROTECTION
- Over Power Protection
- Auto recovery
- Over Voltage Protection
- Auto recovery
- Overt Temperature Protection
- Auto recovery
- Short Circuit Protection
- Protection level 1 (nominal): Continuous, Auto recovery
- Protection level 2 (instantaneous high current): Latch
ISOLATION
- Input-Output
- 4000VAC or 5656VDC
- Input-PE
- 2000VAC or 2828VDC
- Output-PE
- 1500VAC or 2121VDC
ENVIRONMENT
- Operating Temperature
- -30°C...+70°C (with derating)
- Storage Temperature
- -30°C...+85°C
- Temperature Coefficient
- ±0.03%/°C ( 0~50°C )
- ±0.06%/°C ( -30~0°C )
- Altitude During Operation
- 5000m
- Humidity
- 95% RH
- MTBF
- >160,000 h @ 25°C (MIL-HDBK-217F)
- Vibration
- IEC60068-2-6 (10~500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes)
- Shock
- IEC60068-2-27
PHYSICAL
- Dimensions (L x W x H)
- 5.11 x 3.25 x 2.42 Inches ( 129.7 x 82.55 x 61.4 mm ) Tolerance ±0.5 mm
- Weight
- In Progress
SAFETY
- Approval
- UL / IEC / EN 62368 (In Progress)
EMC
- Conducted EMI
- EN55032 Class B (In Progress)
- Radiated EMI
- EN55032 Class A (In Progress)
- EMS
- EN55035 (In Progress)